# I. DC Voltage Measurement with a Digital Multimeter
>[!table] Measured DC Supply Voltages
>
>| Set Voltage [V] | Measured Voltage [V] | Difference [V] | % Error |
>| ---: | ---: | ---: | ---: |
>| 1.0 | 0.999 | -0.001 | 0.100% |
>| -1.0 | -1.001 | -0.001 | 0.100% |
>| 2.0 | 1.999 | -0.001 | 0.050% |
>| -2.0 | -1.999 | 0.001 | 0.050% |
>| 2.5 | 2.499 | -0.001 | 0.040% |
>| -2.5 | -2.499 | 0.001 | 0.040% |
>| 4.0 | 3.999 | -0.001 | 0.025% |
>| -4.0 | -3.999 | 0.001 | 0.025% |
$\text{Mean difference}=-0.00025V$
$\text{Standard deviation}=0.00104V$
The small mean difference and standard deviation indicate that the DMM measurements were precise and closely matched the set PSU voltages.
---
To obtain 35V, the 30V and 5V channels of the PSU were connected in series so that their voltages added together.
$V_{out}=30V+5V=35V$
>[!table] 35V Series Supply Measurement
>
>| Measured Voltage [V] | % Error |
>| ---: | ---: |
>| 35.010 | 0.0286% |
---
>[!table] PSU and DMM Voltage Adjustment Comparison
>
>| Set Voltage [V] | Measured Voltage [V] | % Error |
>| ---: | ---: | ---: |
>| 5.125 | 5.1246 | 0.00780% |
>| 5.126 | 5.1250 | 0.0195% |
Both errors are much less than 5%. The small differences are due to the finite resolution and calibration of the PSU and DMM.
---
Current through a $220\Omega$ resistor with 5V applied:
$I=\frac{V}{R}=\frac{5}{220}=0.0227A=22.7mA$
---
With the PSU set to 5V and a 15mA current limit, the $220\Omega$ resistor would ideally draw 22.7mA. Since this exceeds the current limit, the PSU will enter current-limiting mode and reduce the output voltage.
---
When the $220\Omega$ resistor was connected, the measured voltage dropped to approximately 3.17V because the PSU limited the current to 15mA.
$V=IR=(0.015)(220)=3.30V$
The measured value was close to the expected current-limited voltage.
---
Minimum resistance required to avoid exceeding the 15mA current limit:
$R_{min}=\frac{V}{I}=\frac{5}{0.015}=333.3\Omega$
A resistor of approximately $330\Omega$ was used.
>[!table] Power Supply and DMM Measurements
>
>| Instrument | Voltage [V] | Current [A] |
>| --- | ---: | ---: |
>| Power Supply | 5.000 | 0.015 |
>| DMM | 4.991 | 0.015165 |
The DMM should be relied on for the more precise measurement because it is designed specifically for measurement and provides greater resolution than the PSU display.
# II. Verification of Ohm's Law
>[!table] Measured Voltage and Current for $1k\Omega$ Resistor
>
>| Set Voltage [V] | Measured Current [mA] |
>| ---: | ---: |
>| 0.0 | 0 |
>| 0.5 | 0.5008 |
>| 1.0 | 1.001 |
>| 1.5 | 1.502 |
>| 2.0 | 2.003 |
>| 2.5 | 2.503 |
>| 3.0 | 3.004 |
>| 3.5 | 3.507 |
>| 4.0 | 4.008 |
>| 4.5 | 4.510 |
>| 5.0 | 5.013 |
---
<div style="page-break-after: always;"></div>
>[!figures] I-V Characteristic of $1k\Omega$ Resistor
>![[Pasted image 20260910192104.png|center|500]]
The line of best fit is
$I=1.0024V-0.0014\text{ mA}$
The slope is $1.0024\text{ mA/V}$ and represents the conductance, $\frac{1}{R}$, of the resistor.
$R=\frac{1000}{1.0024}=997.6\Omega$
The linear relationship between current and voltage agrees with Ohm's Law.
---
Measured resistance:
$R_{DMM}=998.06\Omega$
Expected resistance:
$R_{expected}=1000\Omega$
Resistance from the line of best fit:
$R_{fit}=997.6\Omega$
The largest difference is between the expected resistance and the best-fit resistance:
$\Delta R=1000-997.6=2.4\Omega$
$\%\text{ error}=\frac{2.4}{1000}(100)=0.24\%$
This is a reasonable error because it is small and falls within the resistor's specified tolerance.
---
Measured resistance = $998.06\Omega$. The resistor tolerance band is brown, signifying a $\pm1\%$ tolerance.
$R_{min}=1000(0.99)=990\Omega$
$R_{max}=1000(1.01)=1010\Omega$
The measured resistance of $998.06\Omega$ and the best-fit resistance of $997.6\Omega$ both fall within the acceptable range of $990\Omega$ to $1010\Omega$. Therefore, the resistor is within its specified tolerance.